Show simple item record

dc.contributor.author Menezes, Ana Maria Baptista
dc.contributor.author Dumith, Samuel de Carvalho
dc.contributor.author Noal, Ricardo Bica
dc.contributor.author Nunes, Ana Paula
dc.contributor.author Mendonça, Fernanda Ciandrini de
dc.contributor.author Araújo, Cora Luiza Pavin
dc.contributor.author Duval, Marta
dc.contributor.author Caruso, Paulo
dc.contributor.author Hallal, Pedro Rodrigues Curi
dc.date.accessioned 2012-11-25T02:07:31Z
dc.date.available 2012-11-25T02:07:31Z
dc.date.issued 2010
dc.identifier.citation MENEZES, Ana Maria Baptista et al. Validade de um monitor digital de pulso para mensuração de pressão arterial em comparação com um esfigmomanômetro de mercúrio. Arquivos Brasileiros de Cardiologia, v. 94, n. 3, p. 365-370, 2010. Disponível em: <http://www.scielo.br/pdf/abc/v94n3/14.pdf>. Acesso em: 20 nov. 2012. pt_BR
dc.identifier.uri http://repositorio.furg.br/handle/1/2831
dc.description.abstract Fundamento: Medidas válidas da pressão arterial, em situações clínicas e na comunidade, são essenciais para a monitoração dessa variável em nível populacional. Objetivo: Avaliar a validade de um monitor digital de pulso para mensuração da pressão arterial em adolescentes, em comparação com um esfigmomanômetro de mercúrio. Métodos: Um estudo de validação foi realizado na cidade de Pelotas, região sul do Brasil. A pressão arterial foi medida duas vezes, utilizando-se dois esfigmomanômetros diferentes: um aparelho digital de pulso OMRON e um aparelho de mesa BD de mercúrio. Metade da amostra foi medida primeiro através do manômetro digital e depois pelo de mercúrio, enquanto a outra metade foi avaliada na ordem inversa. A concordância entre as duas medidas foi avaliada através do método de Bland & Altman. Resultados: 120 adolescentes com idade entre 14 e 15 anos foram incluídos no estudo (50% de cada sexo). A pressão sistólica média entre os meninos foi de 113,7 mmHg (DP 14,2) usando o manômetro de mercúrio e 115,5 mmHg (DP 15,2) usando o aparelho digital. Os valores equivalentes para a pressão diastólica foram 61,5 mmHg (DP 9,9) e 69,6 mmHg (10,2), respectivamente. Entre as meninas, a pressão sistólica média foi de 104,7 mmHg (DP 10,1) usando o manômetro de mercúrio e 102,4 mmHg (DP 11.9) usando o aparelho digital. Os valores equivalentes para a pressão diastólica foram 60,0 mmHg (DP 10,4) e 65,7 mmHg (DP 7,7), respectivamente. Conclusões: O manômetro digital apresentou alta concordância com o manômetro de mercúrio para medir a pressão arterial sistólica. A concordância foi menor para a pressão arterial diastólica. O uso de equações de correção pode ser uma alternativa para estudos utilizando esse monitor digital de pulso em adolescentes. pt_BR
dc.description.abstract Background: Valid measurements of blood pressure, both at clinical and community settings, are essential for monitoring this variable at the population level. Objective: To evaluate the validity of a wrist digital monitor for measuring blood pressure among adolescents in comparison to a mercury sphygmomanometer. Methods: A validation study was carried out in the city of Pelotas, Southern Brazil. Blood pressure was measured twice using two different sphygmomanometers; an OMRON wrist digital and a desktop BD mercury one. Half of the sample was measured first with the digital manometer and subsequently with the mercury one, whereas the remaining half was evaluated in the opposite order. Agreement between both measures was evaluated using the Bland and Altman method. Results: 120 adolescents aged 14 to 15 years were included (50% of each sex). Mean systolic blood pressure among boys was 113.7 mmHg (SD 14.2) when using the mercury manometer and 115.5 mmHg (SD 15.2) when using the digital one. Equivalent values for diastolic blood pressure were 61.5 mmHg (SD 9.9) and 69.6 mmHg (10.2), respectively. Among girls, the mean systolic blood pressure was 104.7 mmHg (SD 10.1) when using the mercury manometer and 102.4 mmHg (SD 11.9) when using the digital device. Values for diastolic blood pressure were 60.0 mmHg (SD 10.4) and 65.7 mmHg (SD 7.7), respectively. Conclusions: The digital device showed a high level of agreement with the mercury manometer when measuring systolic blood pressure. The level of agreement was lower for diastolic blood pressure. The use of correction equations may be an alternative for studies using this wrist digital monitor in adolescent patients. pt_BR
dc.language.iso por pt_BR
dc.rights open access pt_BR
dc.subject Monitores de pressão arterial pt_BR
dc.subject Determinação da pressão arterial pt_BR
dc.subject Esfigmomanômetros pt_BR
dc.subject Estudos de validação pt_BR
dc.subject Adolescente pt_BR
dc.subject Blood pressure monitors pt_BR
dc.subject Blood pressure determination pt_BR
dc.subject Sphygmomanometers pt_BR
dc.subject Validation studies pt_BR
dc.subject Adolescent pt_BR
dc.title Validade de um monitor digital de pulso para mensuração de pressão arterial em comparação com um esfigmomanômetro de mercúrio pt_BR
dc.title.alternative Validity of a wrist digital monitor for blood pressure measurement in comparison to a mercury sphygmomanometer pt_BR
dc.type article pt_BR


Files in this item

This item appears in the following Collection(s)

:

  • FAMED – Artigos publicados em periódicos
  • Show simple item record