Otimização e validação de método empregando quechers modificado e lc-esi-ms/ms para determinação de agrotóxicos em cebola

Rodrigues, Sherol Acosta; Barbosa, Sergiane Caldas; Furlong, Eliana Badiale; Primel, Ednei Gilberto; Zanella, Renato

Abstract:

An evaluation of the pesticides extraction from onion using a modern sample preparation method (QuEChERS) and determination by liquid chromatography tandem mass spectrometry was carried out. All the calibration curves showed r>0.99. The recoveries ranged between 61.8 and 120.0% with relative standard deviation lower than 20% for all compounds. Due to the occurrence of matrix effect, the quantification was performed using matrix-matched calibration. The limits of quantification of the method were between 0.0005 and 0.05 mg kg-1. The method shows the advantages of not require the clean-up step and consume low volume of organic solvents, decreasing time, costs and residues.

Show full item record

 

Files in this item

This item appears in the following Collection(s)

:

  • EQA - Artigos Publicados em Periódicos